1 September 1990 In-flight validation and calibration of the spectral and radiometric characteristics of the Airborne Visible/Infrared Imaging Spectrometer
Author Affiliations +
Abstract
Through an in-flight calibration experiment at Rogers Dry Lake, California on September 20, 1989, the radiometric and spectral properties of AVIRIS were determined. In-flight spectral channel positions and the spectral response function in 10 regions of the AVIRIS spectral range, taking in all four spectrometers, are shown to agree closely with the corresponding parameters measured in the laboratory. The intraflight stability for the Rogers Dry Lake calibration site is better than 2 percent with the exception of the strong atmospheric water absorptions where the measured radiance is close to zero. This experiment has provided both direct generation of an in-flight spectral and radiometric calibration and validation of the laboratory calibration at the reported level accuracy.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Robert O. Green, Robert O. Green, James E. Conel, James E. Conel, Jack S. Margolis, Jack S. Margolis, Veronique Carrere, Veronique Carrere, Carol J. Bruegge, Carol J. Bruegge, Michael Rast, Michael Rast, Gordon Hoover, Gordon Hoover, } "In-flight validation and calibration of the spectral and radiometric characteristics of the Airborne Visible/Infrared Imaging Spectrometer", Proc. SPIE 1298, Imaging Spectroscopy of the Terrestrial Environment, (1 September 1990); doi: 10.1117/12.21333; https://doi.org/10.1117/12.21333