Paper
1 November 1990 Magnetic resonance image reconstruction from multiple noninteracting receiver coils
Joseph W. Carlson
Author Affiliations +
Abstract
A technique is described for the simultaneous acquisition of NMR data using two independent receiver coils surrounding the same region of tissue which enable the acquisition of all data necessary for an image in a reduced number of acquisitions. This results in a 50% reduction in minimum scan time, but potentially the technique may be extended to multiple receivers with a proportionally higher time savings. The algorithm and imaging procedures are described and example images are described which illustrate the reconstruction. Signal to noise and artifacts are discussed.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Joseph W. Carlson "Magnetic resonance image reconstruction from multiple noninteracting receiver coils", Proc. SPIE 1351, Digital Image Synthesis and Inverse Optics, (1 November 1990); https://doi.org/10.1117/12.23646
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image restoration

Magnetism

Receivers

Back to Top