Paper
1 August 1991 Testing of PZT shifters for interferometric measurements
Joanna Schmit, Tadeusz Piatkowski
Author Affiliations +
Proceedings Volume 1391, Laser Technology III; (1991) https://doi.org/10.1117/12.57157
Event: Laser Technology III, 1990, Szczecin, Poland
Abstract
Nowadays PZT shifters are widely used for interferometric measurements by phase shifting methods. The required accuracy of single step is very high. There for the methods of measuring metrological features of PZT shifters are described specificly intensity methods in Michelson and Sagnac interferometers and Lissajoux figures method 1-D and 2-D Fourier transform method (FTM) in Fizeau interferometer. The 1-D FTM is found most convenient for real time calibration in experimental setup.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Joanna Schmit and Tadeusz Piatkowski "Testing of PZT shifters for interferometric measurements", Proc. SPIE 1391, Laser Technology III, (1 August 1991); https://doi.org/10.1117/12.57157
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Ferroelectric materials

Interferometry

Fizeau interferometers

Mirrors

Fourier transforms

Calibration

Fringe analysis

Back to Top