1 May 1991 Time-resolved transillumination for medical diagnostics
Author Affiliations +
Proceedings Volume 1431, Time-Resolved Spectroscopy and Imaging of Tissues; (1991) https://doi.org/10.1117/12.44183
Event: Optics, Electro-Optics, and Laser Applications in Science and Engineering, 1991, Los Angeles, CA, United States
Abstract
We demonstrate a time-gated technique to reduce the effect of light scattering when transilluminating turbid media such as tissue. The concept is based on transillumination with picosecond laser pulses and time-resolved detection. By detecting only the photons with the shortest travelling time, and thus the least scattered photons, the contrast can be enhanced. Measurements on a tissue phantom as well as breast tissue in vitro are presented. It is demonstrated that differences in scattering properties may be more pronounced than differences in absorption properties when demarcating tumor from normal tissue.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Roger Berg, Roger Berg, Stefan Andersson-Engels, Stefan Andersson-Engels, Olof Jarlman, Olof Jarlman, Sune Svanberg, Sune Svanberg, } "Time-resolved transillumination for medical diagnostics", Proc. SPIE 1431, Time-Resolved Spectroscopy and Imaging of Tissues, (1 May 1991); doi: 10.1117/12.44183; https://doi.org/10.1117/12.44183