1 May 1991 Measurement of peroxyacetyl nitrate, NO2, and NOx by using a gas chromatograph with a luminol-based detector
John W. Drummond, Gervase I. Mackay, Harold I. Schiff
Author Affiliations +
Proceedings Volume 1433, Measurement of Atmospheric Gases; (1991) https://doi.org/10.1117/12.46168
Event: Optics, Electro-Optics, and Laser Applications in Science and Engineering, 1991, Los Angeles, CA, United States
Abstract
A commercial instrument for measuring PAN (peroxyacetyl nitrate), NO2, and optionally, NOx, is described. It is based on a gas chromatograph (GC), and uses a LuminoxR luminol detector following conversion of the PAN to NO2. A detection limit (based on a peak height of 3 X noise) of 30 pptV is achieved for injections made every five minutes. The instrument is computer driven and contains its own peak integrator and graphics output. No interferences have been identified, including chlorinated hydrocarbons, NO2 and ozone. However, the PAN analog peroxypropanyl nitrate (PPN) is not fully resolved from PAN. The maintenance interval is monthly. The calibration of the instruments is relatively easy: known mixing ratios of NO2 in air are used to calibrate the sample injection system and LuminoxR detector. The passage of PAN through the column and the subsequent conversion of PAN to NO2 has been found to be stable over times of one year, and is reproducible between instruments. Atmospheric measurements for PAN have been carried out for both very polluted environments and very clean environments. The results from a month long formal intercomparison between the LuminoxR PAN and two GCs equipped with electron capture detectors is presented. Those results demonstrate that the luminol based PAN detector can yield accurate measurements of ambient PAN concentrations. A method for calibrating PAN instruments in the field is described.
© (1991) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
John W. Drummond, Gervase I. Mackay, and Harold I. Schiff "Measurement of peroxyacetyl nitrate, NO2, and NOx by using a gas chromatograph with a luminol-based detector", Proc. SPIE 1433, Measurement of Atmospheric Gases, (1 May 1991); https://doi.org/10.1117/12.46168
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

NOx

Sensors

Nitrous oxide

Prototyping

Analog electronics

Environmental sensing

Back to Top