1 April 1992 Contactless measurement of the spectral emissivity and temperature of surfaces with a Fourier transform infrared spectrometer
Author Affiliations +
Abstract
A method for the measurement of the spectral emissivity and temperature of a surface is presented. The spectrum of the infrared radiation of a heated surface is measured by means of a Fourier transform infrared spectrometer. To obtain emission spectra in units of spectral radiance a special calibration procedure is described which uses three black bodies as reference sources thus eliminating the need to measure the black bodies' surface temperatures. A theoretical emission spectrum of the investigated surface's radiance is fitted to the calibrated measurement resulting in (in the Gaussian sense) optimal values of spectral emissivity and temperature of the surface.
© (1992) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Erwin Lindermeir, Volker Tank, Peter Haschberger, "Contactless measurement of the spectral emissivity and temperature of surfaces with a Fourier transform infrared spectrometer", Proc. SPIE 1682, Thermosense XIV: An Intl Conf on Thermal Sensing and Imaging Diagnostic Applications, (1 April 1992); doi: 10.1117/12.58553; https://doi.org/10.1117/12.58553