3 February 1993 Energy response of astronomical CCD x-ray detectors
Author Affiliations +
Abstract
We present a description of Monte Carlo simulations of the pulse height energy response of CCD x-ray detectors. Effects of event charge splitting between pixels are accounted for, and the variation of energy resolution and detection efficiency with event selection criteria are discussed. We show that there are important implications for background rejection efficiency, on-board calibration, and ground-based data reduction. In orbit energy resolution may become degraded by traps created by radiation damage. We present an analysis of trapping and emission time scales, which allows us to predict the energy resolution for a wide range of device operating conditions.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
David H. Lumb, John A. Nousek, "Energy response of astronomical CCD x-ray detectors", Proc. SPIE 1736, X-Ray Detector Physics and Applications, (3 February 1993); doi: 10.1117/12.140471; https://doi.org/10.1117/12.140471
PROCEEDINGS
11 PAGES


SHARE
RELATED CONTENT

Junction CCD-based x-ray sensor for dental applications
Proceedings of SPIE (September 14 1993)
Particle background in CCD detectors
Proceedings of SPIE (November 19 1993)
MOS CCDs for the European Photon Imaging Camera
Proceedings of SPIE (November 19 1993)