30 April 1993 Dissolved-oxygen quenching of in-situ fluorescence measurements
Author Affiliations +
Abstract
In-situ fluorescence measurements of aromatic organic ground water contaminants do not always agree with gas chromatographic methods. Dissolved oxygen quenching of fluorescence may be an interferant in field measurements. Two standard fluorescent aromatics, quinine sulfate and naphthalene, were evaluated in this study. Over the range of dissolved oxygen concentrations expected to be encountered in the field, no effects of oxygen quenching on fluorescence of these compounds was observed. Quenching of quinine sulfate fluorescence by sodium chloride was observed using this system. Sodium chloride quenching was shown to follow the Stern-Volmer relation.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Wayne Chudyk, David Tonaszuck, and Kenneth Pohlig "Dissolved-oxygen quenching of in-situ fluorescence measurements", Proc. SPIE 1796, Chemical, Biochemical, and Environmental Fiber Sensors IV, (30 April 1993); doi: 10.1117/12.143530; https://doi.org/10.1117/12.143530