30 April 1993 Spectroscopic gas sensing with infrared hollow waveguides
Author Affiliations +
Abstract
Spectroscopic gas sensing was carried out by using an infrared hollow waveguide as a capillary flow cell. A ZnS-coated Ag hollow waveguide is generally the most suitable selection for use as a flow cell, since it exhibits high transmittance over the wide spectral range around 10 micrometers wavelength. For the corrosive gases that cause serious damage to the ZnS film, SiO2- or GeO2-based glass hollow waveguides are applicable. With these low-loss hollow waveguides, CH4, n-C4H10, NO2, and SO2 gases were measured successfully, and a fast response in gas detection as well as a remarkable reduction of gas consumption was demonstrated.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mitsunori Saito, Mitsunori Saito, Shin-ya Sato, Shin-ya Sato, Mitsunobu Miyagi, Mitsunobu Miyagi, } "Spectroscopic gas sensing with infrared hollow waveguides", Proc. SPIE 1796, Chemical, Biochemical, and Environmental Fiber Sensors IV, (30 April 1993); doi: 10.1117/12.143526; https://doi.org/10.1117/12.143526