24 June 1993 Technical monograph of the True VistaTM bifocal intraocular lens
Author Affiliations +
Proceedings Volume 1877, Ophthalmic Technologies III; (1993) https://doi.org/10.1117/12.147525
Event: OE/LASE'93: Optics, Electro-Optics, and Laser Applications in Scienceand Engineering, 1993, Los Angeles, CA, United States
Abstract
The STORZTM TRUE VISTATM bifocal intraocular lens was evaluated in a laboratory setting and during clinical trials. The modulation transfer function and through focus response were measured in a simulated eye chamber for selection of optimal add zone size, location, power, and pupil dependency. Results indicate the optimal design to be an annular add power of +4 D having a 1.5 millimeter inner diameter and a 2.6 millimeter outer diameter. Optical performance was predicted using computer ray trace and Fourier analysis. Laboratory testing confirmed design performance achieved through the manufacturing process. Clinical results correlate well with through focus response but less well with modulation transfer function.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
William J. Neubert, Ta-Chung Hsia, George F. Green, Mark S. Jones, Brian D. McCary, Susan M. Kastigar, and Jose H. Bedoya "Technical monograph of the True VistaTM bifocal intraocular lens", Proc. SPIE 1877, Ophthalmic Technologies III, (24 June 1993); https://doi.org/10.1117/12.147525