14 September 1993 Cd1-xZnxTe detector imaging array
Author Affiliations +
Abstract
A prototype portable gamma ray camera using 32 X 32 channels was developed. An experimental 3 X 3 sub-array of 5 mm X 5 mm CZT detectors was fabricated for use in system checkout and to investigate the applicability of CZT imaging arrays to nuclear medical imaging. Experiments were carried out to make a direct comparison of the imaging capabilities of the CZT sub-array with a state-of-the-art Anger camera. In a linespread study using a Tc-99m source embedded in a tissue equivalent absorber, contrasts of 9.5 for the CZT array and 3.4 for the Anger camera were observed. In a dynamic imaging experiment, the CZT imager appeared to have comparable resolution to and be somewhat more regular than the Anger camera.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jack F. Butler, Stan J. Friesenhahn, Clinton L. Lingren, Boris A. Apotovsky, F. Patrick Doty, William L. Ashburn, and William P. Dillon "Cd1-xZnxTe detector imaging array", Proc. SPIE 1896, Medical Imaging 1993: Physics of Medical Imaging, (14 September 1993); https://doi.org/10.1117/12.154607