14 September 1993 Development of a dual-energy kinestatic charge detector
Author Affiliations +
Abstract
A design for a dual-energy Kinestatic Charge Detector (KCD) with segmented charge collection fingers is discussed. The front segment of the KCD charge collectors will produce a digital low-energy image and the back segments will produce a digital high-energy image. A gap between the front and back collectors acts as a mid filter to increase the separation between the mean energies absorbed in the front and back segments. This dual-energy KCD design was optimized by using computer simulations to maximize a figure of merit defined as the square of the signal-to-noise ratio divided by the average absorbed dose. The optimal KCD parameters are presented for a dual-energy KCD designed for mammography. Preliminary experimental data are given for a small non imaging dual-energy KCD research detector.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Robert J. Endorf, Sunil Kulatunga, David C. Spelic, Frank A. DiBianca, Herbert D. Zeman, "Development of a dual-energy kinestatic charge detector", Proc. SPIE 1896, Medical Imaging 1993: Physics of Medical Imaging, (14 September 1993); doi: 10.1117/12.154590; https://doi.org/10.1117/12.154590