19 November 1993 Berkeley EUV spectrometer microchannel plate detectors for ORFEUS
Author Affiliations +
Abstract
We have developed detectors design for high resolution spectroscopic imaging for the EUV spectrometer on the ORFEUS-SPAS mission. The detectors employ spherically curved microchannel plates and a delay line read out system. We present results from the testing and calibration of the detectors prior to their integration into the spectrometer. The design, MCP preconditioning, and imaging characteristics are discussed for 2 rectangular spectral detectors and for the fine guidance detector. Each spectral detector achieves a 30 X 100 micrometers FWHM resolution element size over a 95 mm X 25 mm anode format with good linearity (< 30 micrometers ) in the spectral direction. The fine guidance detector, used for drift corrections, achieves good resolution and will meet the necessary centroiding requirements.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Joseph M. Stock, Oswald H. W. Siegmund, Mark Hurwitz, Richard Raffanti, C. Stuart Bowyer, and Michael L. Lampton "Berkeley EUV spectrometer microchannel plate detectors for ORFEUS", Proc. SPIE 2006, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy IV, (19 November 1993); doi: 10.1117/12.162826; https://doi.org/10.1117/12.162826