Paper
16 December 1993 CCD-based detector for crystallographic applications using laboratory x-ray sources
Walter Charles Phillips, Martin J. Stanton, Daniel M. O'Mara, Youli Li, Istvan Naday, Edwin M. Westbrook
Author Affiliations +
Abstract
We describe a modular detector designed primarily for macro-molecular x-ray crystallography experiments which use a rotating-anode x-ray source. The module consists of a fiber optic demagnifying taper with a phosphor x-ray converter deposited on the large end of the taper and a CCD bonded to the small end. Two or more modules can be used together to increase the total detector area. The detector components are chosen to optimize data-collection efficiency while keeping the overall cost relatively low and the reliability high. Performance measurements for a prototype detector of this design are presented.
© (1993) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Walter Charles Phillips, Martin J. Stanton, Daniel M. O'Mara, Youli Li, Istvan Naday, and Edwin M. Westbrook "CCD-based detector for crystallographic applications using laboratory x-ray sources", Proc. SPIE 2009, X-Ray Detector Physics and Applications II, (16 December 1993); https://doi.org/10.1117/12.164732
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Charge-coupled devices

X-rays

CCD image sensors

X-ray detectors

Fiber optics

CCD cameras

RELATED CONTENT

CCD-based detector for x-ray crystallography
Proceedings of SPIE (July 12 1994)
Modular CCD detector for x-ray crystallography
Proceedings of SPIE (July 12 1993)
Operational experience of a large area x ray camera for...
Proceedings of SPIE (October 06 1999)
X-ray detector for time-resolved studies
Proceedings of SPIE (November 01 1991)

Back to Top