23 June 1994 Comparison of the EyeSys corneal analysis system and the TMS topographic modeling system using a bicurve test surface
Cynthia J. Roberts
Author Affiliations +
Proceedings Volume 2126, Ophthalmic Technologies IV; (1994) https://doi.org/10.1117/12.178582
Event: OE/LASE '94, 1994, Los Angeles, CA, United States
Abstract
The EyeSys Corneal Analysis System and the TMS Topographic Modeling System were compared, both when properly aligned and under specified conditions of misalignment, in their ability to map a sharp transition in curvature in order to model the type of surface topography that might be produced after photorefractive keratectomy. Results indicate that neither system can accurately reconstruct the transition zone, and that the source of the error is data averaging across the meridians to produce a smooth display for the EyeSys system, and lesser averaging combined with a fundamental error in reconstructing the peripheral region for the TMS system. In addition, the TMS is shown to possess greater sensitivity to errors of decentration and defocus for the bicurve test surface, while the EyeSys system is shown to be relatively immune to similar conditions of misalignment.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Cynthia J. Roberts "Comparison of the EyeSys corneal analysis system and the TMS topographic modeling system using a bicurve test surface", Proc. SPIE 2126, Ophthalmic Technologies IV, (23 June 1994); https://doi.org/10.1117/12.178582
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Eye models

Systems modeling

Spherical lenses

Data storage

Computing systems

Cornea

Data acquisition

Back to Top