7 September 1994 Novel CO2 laser system for hard tissue ablation
Author Affiliations +
Abstract
A line-tunable and variable pulselength CO2 laser system for ablation of dental and bone tissues has been tested. To avoid the undesirable effects of plasma formation, acoustic shock waves and collateral thermal damage to the tissues, we have found an optimum laser pulselength in the 10 - 20 microseconds range. At a fluence of 30 J/cm2 and laser wavelength of 10.6 micrometers the ablation depths in enamel (dentin) were 12(18) micrometers and no significant noise was observed. Without water cooling enamel showed no visible color change, while some minimal carbonization of dentin was present even at low pulse repetition frequencies.
© (1994) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Vladimir B. Krapchev, Vladimir B. Krapchev, Christopher D. Rabii, Christopher D. Rabii, James A. Harrington, James A. Harrington, } "Novel CO2 laser system for hard tissue ablation", Proc. SPIE 2128, Laser Surgery: Advanced Characterization, Therapeutics, and Systems IV, (7 September 1994); doi: 10.1117/12.184916; https://doi.org/10.1117/12.184916