30 May 1995 Development of a new clinical NIRS sensor for absolute quantification of cerebral oxygenation
Author Affiliations +
Abstract
A breadboard NIR sensor system has been developed based on the novel Psuedo-Random Modulation/Resolution Enhancement Technique (PRM/RET) and using a medium power diode laser system. The breadboard sensor system has been used to measure the photon migration in a phantom model consisting of milk and additive inks. The result clearly demostrates the unique capability of the PRM/RET sensor system for high-resolution time-resolved spectroscopic measurement with a nanosecond resolution which is currently limited by the digitizer bandwidth. The data obtained from this experiment was analyzed to retrieve the optical parameters of the phantom quantitatively. Based on the performance of the breadboard system, a prototype system configuration is being designed to obtain over a few GHz bandwidth and a few tens of psec precision which will be applied for the simple and accurate retrieval of the cerebral optical properties.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hyo Sang Lee, L. Zheng, Anthony Notari, "Development of a new clinical NIRS sensor for absolute quantification of cerebral oxygenation", Proc. SPIE 2389, Optical Tomography, Photon Migration, and Spectroscopy of Tissue and Model Media: Theory, Human Studies, and Instrumentation, (30 May 1995); doi: 10.1117/12.210006; https://doi.org/10.1117/12.210006
PROCEEDINGS
10 PAGES


SHARE
RELATED CONTENT

Coherent sources for mid-infrared laser spectroscopy
Proceedings of SPIE (December 23 2016)
Atmospheric transmission for cesium DPAL using TDLAS
Proceedings of SPIE (February 24 2012)
Fiber optic systems for remote laser spectroscopy
Proceedings of SPIE (April 28 1999)
Modulation cancellation method for laser spectroscopy
Proceedings of SPIE (January 24 2011)

Back to Top