22 May 1995 Holmium:YAG thermokeratoplasty: treatment parameters for the correction of astigmatism based upon enucleated human eyes using an application mask
Author Affiliations +
Abstract
The refractive outcome of thermokeratoplasty depends upon the location and angle of the coagulation spots, applied with a focusing handpiece onto the corneal surface. Accuracy can be enhanced using a specially designed application mask. An astigmatism correction was performed on 10 human donor eyes (Holmium 25, Technomed, FRG, 15 Hz, 20 mJ/pulse, 25 pulses) with an optical zone of 8.1 mm, 5 eyes received a free hand laser application (marked positions) and the other 5 eyes were treated using a suctioned metal mask with drills for the handpiece (optical zone 8.1 mm). To compare the results a silicone replica was taken and analyzed by a confocal laser microtopometer. The refractive change for the steepest meridian was 10 D with a standard deviation of +/- 3.7 D for the free hand application. Using the application mask the refractive outcome was 9.8 D with a standard deviation of only 0.8 D. Using the application mask the standard deviation for the induced refractive change decreases by a factor of five.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Martin Kriegerowski, Katja Rassmann, Theo Oltrup, Thomas Bende, Benedikt J. Jean, "Holmium:YAG thermokeratoplasty: treatment parameters for the correction of astigmatism based upon enucleated human eyes using an application mask", Proc. SPIE 2393, Ophthalmic Technologies V, (22 May 1995); doi: 10.1117/12.209827; https://doi.org/10.1117/12.209827