8 May 1995 Development of large field of view and high-definition real-time digital radiography system
Author Affiliations +
Abstract
We have developed a high definition real-time DR system capable of 2000 line scan adopting a high resolution TV camera using a 2' Saticon imaging tube and a 16' high definition x-ray image intensifier. As a result, we obtained the dynamic range of a DR image, namely latitude of x rays, of 79 that is comparable to that of the film-screen system, and a limiting spatial resolution of 2.5 lp/mm for a maximum field-of-view of 16'.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hisatake Yokouchi, Hisatake Yokouchi, Fumitaka Takahashi, Fumitaka Takahashi, Yoichi Onodera, Yoichi Onodera, Ken Ueda, Ken Ueda, Tetsuro Endo, Tetsuro Endo, } "Development of large field of view and high-definition real-time digital radiography system", Proc. SPIE 2432, Medical Imaging 1995: Physics of Medical Imaging, (8 May 1995); doi: 10.1117/12.208368; https://doi.org/10.1117/12.208368