8 May 1995 Imaging performance of a large-field kinestatic charge detector for digital radiography
Author Affiliations +
Abstract
The initial performance of a digital radiographic system incorporating a large-field (2016- channel) kinestatic charge detector and data acquisition electronics is discussed. The measured modulation transfer function of the system is 20% at 4 cy/mm. The measured detective quantum efficiency is 40 - 60%. These results are comparable with or better than those of current clinical (rare-earth film-screen and storage phosphor) systems. First images from the large-field system are shown and compared with those from commercial systems. Future system improvements in process or in planning are discussed.
© (1995) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Frank A. DiBianca, Frank A. DiBianca, Sreenivas Devidas, Sreenivas Devidas, George C. Giakos, George C. Giakos, Shashidhar Kollipara, Shashidhar Kollipara, Joseph S. Laughter, Joseph S. Laughter, Azad Mahmud, Azad Mahmud, Senthilkumar Nagarajan, Senthilkumar Nagarajan, Qian Peng, Qian Peng, Carlos Rodriguez, Carlos Rodriguez, Herbert D. Zeman, Herbert D. Zeman, "Imaging performance of a large-field kinestatic charge detector for digital radiography", Proc. SPIE 2432, Medical Imaging 1995: Physics of Medical Imaging, (8 May 1995); doi: 10.1117/12.208359; https://doi.org/10.1117/12.208359