19 January 1996 Endotracheal tubes for laser surgery: temperature rise and clinical experiences
Author Affiliations +
Abstract
An infrared camera was used to measure the temperature rise which takes place in endotracheal tubes under laser irradiation. It was seen that a metallic tube was heated up within a second to temperatures of 200 degrees to 300 degrees Celsius which was very destructive to the PVC conduits inside of the tube. A compound tube, on the other hand, reached temperatures of only 38 degrees Celsius at its inner surface. The thermal induced destruction of the conduits inside of the metallic tube is seen as the reason for complications like airway blocking. Furthermore preliminary results of a randomized clinical study are presented, showing that the metallic tube needed higher pressure levels than the compound tube.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Hans-Jochen Foth, Hans-Jochen Foth, Dirk H. Meyer, Dirk H. Meyer, Antonio Baker-Schreyer, Antonio Baker-Schreyer, Wolfgang Bergler, Wolfgang Bergler, Karl Hoermann, Karl Hoermann, Joseph Umgemach, Joseph Umgemach, } "Endotracheal tubes for laser surgery: temperature rise and clinical experiences", Proc. SPIE 2623, Medical Applications of Lasers III, (19 January 1996); doi: 10.1117/12.230367; https://doi.org/10.1117/12.230367