You have requested a machine translation of selected content from our databases. This functionality is provided solely for your convenience and is in no way intended to replace human translation. Neither SPIE nor the owners and publishers of the content make, and they explicitly disclaim, any express or implied representations or warranties of any kind, including, without limitation, representations and warranties as to the functionality of the translation feature or the accuracy or completeness of the translations.
Translations are not retained in our system. Your use of this feature and the translations is subject to all use restrictions contained in the Terms and Conditions of Use of the SPIE website.
18 December 1995Coherence effects in modeling laser Doppler perfusion flowmetry
When performing direct-contact laser-Doppler flowmetry on experimental flow models, the power spectra of the detector signal can be obtained by homodyne or by heterodyne detection. Especially with uniformly moving probe particles coherence effects are observed, leading to changes in the width of the power spectrum. Due to destructive interference effects at the detector surface, in the measured homodyne spectra the contribution of relatively high Doppler frequencies is suppressed compared with that of lower frequencies. This results in spectra which are narrower than expected theoretically. This effect allows us to investigate the importance of the relative amount of coherence areas at the surface of the detector.
The alert did not successfully save. Please try again later.
Frits F. M. de Mul, Wiendelt Steenbergen, T. Vonck, Jan Greve, "Coherence effects in modeling laser-Doppler perfusion flowmetry," Proc. SPIE 2626, Photon Propagation in Tissues, (18 December 1995); https://doi.org/10.1117/12.228677