7 May 1997 Accuracy of distance measurements in biplane angiography
Klaus D. Toennies, Satoru Oishi, David Koster, Gerhard Schroth
Author Affiliations +
Abstract
Distance measurements of the vascular system of the brain can be derived from biplanar digital subtraction angiography (2p-DSA). The measurements are used for planning of minimal invasive surgical procedures. Our 90 degree-fixed-angle G- ring angiography system has the potential of acquiring pairs of such images with high geometric accuracy. The sizes of vessels and aneurysms are estimated applying a fast and accurate extraction method in order to select an appropriate surgical strategy. Distance computation from 2p-DSA is carried out in three steps. First, the boundary of the structure to be measured is detected based on zero-crossings and closeness to user-specified end points. Subsequently, the 3D location of the center of the structure is computed from the centers of gravity of its two projections. This location is used to reverse the magnification factor caused by the cone-shaped projection of the x-rays. Since exact measurements of possibly very small structures are crucial to the usefulness in surgical planning, we identified mechanical and computational influences on the geometry which may have an impact on the measurement accuracy. A study with phantoms is presented distinguishing between the different effects and enabling the computation of an optimal overall exactness. Comparing this optimum with results of distance measurements on phantoms whose exact size and shape is known, we found, that the measurement error for structures of size of 20 mm was less than 0.05 mm on average and 0.50 mm at maximum. The maximum achievable accuracy of 0.15 mm was in most cases exceeded by less than 0.15 mm. This accuracy surpasses by far the requirements for the above mentioned surgery application. The mechanic accuracy of the fixed-angle biplanar system meets the requirements for computing a 3D reconstruction of the small vessels of the brain. It also indicates, that simple measurements will be possible on systems being less accurate.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Klaus D. Toennies, Satoru Oishi, David Koster, and Gerhard Schroth "Accuracy of distance measurements in biplane angiography", Proc. SPIE 3031, Medical Imaging 1997: Image Display, (7 May 1997); https://doi.org/10.1117/12.273907
Lens.org Logo
CITATIONS
Cited by 11 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Distortion

Image intensifiers

Angiography

3D image processing

Distance measurement

X-ray sources

Imaging systems

Back to Top