23 May 1997 Calibration of an open-path FTIR spectrometer for methane, ethylene, and carbon monoxide using a fixed 20-m multipass cell
Author Affiliations +
Abstract
A calibration experiment was performed by the Fachhochschule Duesseldorf in cooperation with the official calibration laboratory of the Environmental State Agency of Northrhine Westphalia. This was the first calibration of an OP-FTIR spectrometer at an official calibration laboratory in Germany. The calibration measurements were carried out along the standard ISO/DIS 9169, 'air quality - determination of performance characteristics of measurement methods'. A multipass cell with a light path of 20m was used. During these measurements the deviation between the calibration gas concentration and the measured concentration increased rapidly with higher concentration. This indicates a nonlinear calibration function of the OP-FTIR spectrometer used here and might be addressed to the used spectral resolution of 1.0 wavenumbers. The calibration measurements are still running for other gases and other OP-FTIR spectrometers.
© (1997) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Torsten Lamp, M. Radmacher, Konradin Weber, Andrea Gaertner, R. Nitz, and G. Broeker "Calibration of an open-path FTIR spectrometer for methane, ethylene, and carbon monoxide using a fixed 20-m multipass cell", Proc. SPIE 3107, Remote Sensing of Vegetation and Water, and Standardization of Remote Sensing Methods, (23 May 1997); doi: 10.1117/12.274748; https://doi.org/10.1117/12.274748