Paper
1 May 1998 Single-crystal PZN/PT transducers
Patrick D. Lopath, Seung Eek Eagle Park, K. Kirk Shung, Thomas R. Shrout
Author Affiliations +
Abstract
Single crystal Pb(Zn113Nb2i3)03 (PZN) I PbTiO3 (PT) solid solutions have been investigated as the active materials in medical diagnostic transducers. The compositionally controlled properties of this crystal system allow the application of this material to transducers which span the frequency range of modern diagnostic transducers. Phased array elements operating at 5MHz were constructed from rhombohedral PZN I 4.5% PT single crystals and compared to PZT 5-H elements. The single crystal element displayed similar sensitivity over a broader bandwidth response. Transducer construction techniques employed differ from standard ceramic arrays. Crystallographic alignment was found to be a significant contributor to piezoelectric performance. The crystal structure and phase diagram play a major role in processing steps requiring heating such as poling and electroding. Emphasizing the versatility of the PZN/PT single crystals, a 20 MHz single element was also constructed, using tetragonal PZN /12% PT. Its response was compared to theoretical pulseechoes generated via the KLM model.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Patrick D. Lopath, Seung Eek Eagle Park, K. Kirk Shung, and Thomas R. Shrout "Single-crystal PZN/PT transducers", Proc. SPIE 3341, Medical Imaging 1998: Ultrasonic Transducer Engineering, (1 May 1998); https://doi.org/10.1117/12.308007
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Transducers

Crystals

Chemical elements

Dielectrics

Ferroelectric materials

Ceramics

Electrodes

Back to Top