Paper
19 November 1998 X-ray fluorescence analysis based on Kumakhov optics
Svetlana V. Nikitina, Nariman S. Ibraimov, Alexander S. Stcherbakov
Author Affiliations +
Abstract
The use of Kumakhov optics in x-ray fluorescence analysis is considered. Thanks to high efficiency of a polycapillary lens the concentration of x-ray detector in a close proximity to the sample decreases sufficiently the time of exposure. It is shown experimentally that in the case of use of a small x-ray source with power of 2W the minimal detection limit may be of the order of 0.1 pg. A new portable x-ray fluorescence analyzer based on Kumakhov optics is described. Minimal detection limit may achieve 10-16 g if 100 W microfocus source and the lens with approximately 10 micrometers focal spot are used.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Svetlana V. Nikitina, Nariman S. Ibraimov, and Alexander S. Stcherbakov "X-ray fluorescence analysis based on Kumakhov optics", Proc. SPIE 3444, X-Ray Optics, Instruments, and Missions, (19 November 1998); https://doi.org/10.1117/12.331260