22 December 1998 Calibration method for radiometric and wavelength calibration of a spectrometer
Author Affiliations +
Abstract
A new calibration target or Certified Reference Material (CRM) has been designed that uses violet, orange, green and cyan dyes ont cotton paper. This paper type was chosen because it has a relatively flat spectral response from 400 nm to 700 nm and good keeping properties. These specific dyes were chosen because the difference signal between the orange, cyan, green and purple dyes have certain characteristics that then a low the calibration of an instrument. The ratio between the difference readings is a direct function of the center wavelength of a given spectral band. Therefore, the radiometric and spectral calibration can be determined simultaneously from the physical properties of the reference materials.
© (1998) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Edward M. Granger, "Calibration method for radiometric and wavelength calibration of a spectrometer", Proc. SPIE 3648, Color Imaging: Device-Independent Color, Color Hardcopy, and Graphic Arts IV, (22 December 1998); doi: 10.1117/12.334562; https://doi.org/10.1117/12.334562