28 May 1999 Operation of amorphous silicon detectors for chest radiography within system latency requirements
Author Affiliations +
Abstract
A timing sequence for operation of an amorphous silicon detector within a digital chest radiography system is evaluated. Measurements of large-area spatial non-uniformities (i.e., artifacts) resulting from various implementations of the sequence are presented. Based on these measurements, a method for operation is determined which successfully eliminates large-area artifacts and which satisfies several important system latency requirements. In addition, a method for quantifying artifacts of the type observed in these experiments is presented.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
John M. Boudry, John M. Boudry, } "Operation of amorphous silicon detectors for chest radiography within system latency requirements", Proc. SPIE 3659, Medical Imaging 1999: Physics of Medical Imaging, (28 May 1999); doi: 10.1117/12.349508; https://doi.org/10.1117/12.349508