28 May 1999 Variations in the geometrical distortion of x-ray image intensifiers
Author Affiliations +
Abstract
The paper investigates the variability of the geometrical deformation in digital X-ray image intensifier images as a function of acquisition parameters and time. We have increased our understanding of the deformation patterns and of how they are influenced by normal clinical use. Our goal is to reduce the frequency of calibration and the complexity of the protocol for diagnostic and interventional procedures so that geometrical distortion correction can become part of daily clinical practice.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Bert Verdonck, Philippe Bourel, Eric Coste, Frans A. Gerritsen, Jean Rousseau, "Variations in the geometrical distortion of x-ray image intensifiers", Proc. SPIE 3659, Medical Imaging 1999: Physics of Medical Imaging, (28 May 1999); doi: 10.1117/12.349501; https://doi.org/10.1117/12.349501
PROCEEDINGS
10 PAGES


SHARE
Back to Top