Paper
1 November 1982 New Dual Rectangular Photomultiplier Tube For Positron CT.
T. Yamashita, M. Ito, T. Hayashi
Author Affiliations +
Abstract
A new dual rectangular photomultiplier tube has been developed for the positron CT. It has two independent segments in a single 24mm square glass envelope. The coincidence time resolution with BG0 scintillators equal to 2.9ns in FWHM and 6.2ns in FWTM was obtained for 511 KeV positron annihilation ,γ-rays. The experimental results promise the possibility of a new PCT system design-ing of a high packing ratio and spatial resolution with good optical coupling to small rectangular BGO scintillators.
© (1982) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
T. Yamashita, M. Ito, and T. Hayashi "New Dual Rectangular Photomultiplier Tube For Positron CT.", Proc. SPIE 0372, Physics and Engineering in Medical Imaging, (1 November 1982); https://doi.org/10.1117/12.934517
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Scintillators

Photomultipliers

Glasses

Quantum efficiency

Spatial resolution

Electrodes

Sensors

Back to Top