9 December 1999 Fast-response sensors for dissolved oxygen
Author Affiliations +
Abstract
In this work we have been developing fast response optical sensors for dissolved oxygen determination based on measuring the luminescence lifetime of a metalloporphyrin polymer. The sensor is produced by electropolymerization of the metalloporphyrin monomer units of platinum tetraphenylporphyrin (Pt-TPP), platinum octaethylporphyrin (Pt-OEP), palladium tetraphenylporphyrin (Pd-TPP) or palladium octaethylporphyrin (Pd-OEP). The polymerization process results in films which are in the region of micrometers thick. The Stern-Volmer quenching constants as determined from luminescence lifetime measurements for these sensors range from 0.90 (mg 1-1)-1) for Pt-TPP to 1.83 (mg 1-1)-1) for Pd-OEP. The response time of these sensors to a step change from an oxygen free to an oxygen saturated solution is in the millisecond region.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Sheila Smith, Sheila Smith, Alan Hamill, Alan Hamill, Mahesh Uttamlal, Mahesh Uttamlal, Michael Campbell, Michael Campbell, David Bailly, David Bailly, } "Fast-response sensors for dissolved oxygen", Proc. SPIE 3860, Fiber Optic Sensor Technology and Applications, (9 December 1999); doi: 10.1117/12.372962; https://doi.org/10.1117/12.372962