6 June 2000 Automated 3D-stent localization from intravascular ultrasound image sequences
Author Affiliations +
Abstract
This paper presents a new method for automated detection of stents. The method consists of three sequential steps. At first, a 3D to 2D projection procedure is applied to find the global stent location. A local search strategy was designed, using a combination of characteristic image features to extract the sampling point candidates of the stent in each of the original cross-sectional images. Finally, the preselected points are accepted or rejected depending on a set of a priori criteria for position and shape. Using the resulting stent points, geometrical parameters of the stent are automatically calculated and a wire frame model is generated for 3D surface reconstruction. Thus, in combination with our algorithms for automated detection of the lumen cross-sectional area, the new method is an essential component for 3D visualization of stents and the automated quantification of the degree of in- stent restenosis. The evaluation is based on in vitro and in vivo recordings. The results show that the new algorithm is well-suited to replace time consuming manual segmentation and measurements.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Michael Schmauder, Michael Schmauder, Steffen Zeiler, Steffen Zeiler, C. M. Gross, C. M. Gross, Juergen Waigand, Juergen Waigand, Reinhold Orglmeister, Reinhold Orglmeister, } "Automated 3D-stent localization from intravascular ultrasound image sequences", Proc. SPIE 3979, Medical Imaging 2000: Image Processing, (6 June 2000); doi: 10.1117/12.387656; https://doi.org/10.1117/12.387656
PROCEEDINGS
9 PAGES


SHARE
Back to Top