Paper
21 July 2000 Laser-produced plasma-based reflectometer for EUV metrology
Stanley Mrowka, James H. Underwood, Eric M. Gullikson, Phillip J. Batson
Author Affiliations +
Abstract
An EUV/soft x-ray reflectometer is described which is based on a laser produced plasma source. Continuous wavelength scans in the range 4 nm to 40 nm are possible using a spherical grating monochromator. Focusing optics enable sub-millimeter illumination spot sizes to be achieved at the sample. Rotation and translation stages allow measurements to be made from grazing incidence to 85 degrees on samples as large as 200 mm diameter.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Stanley Mrowka, James H. Underwood, Eric M. Gullikson, and Phillip J. Batson "Laser-produced plasma-based reflectometer for EUV metrology", Proc. SPIE 3997, Emerging Lithographic Technologies IV, (21 July 2000); https://doi.org/10.1117/12.390124