16 August 2000 First test results and calibration methods of CONICA as a stand-alone device
Author Affiliations +
Abstract
First laboratory test result of CONICA are presented for the variety of observation modes: using the final ALADDIN- Detector, IR images in direct, spectroscopic and polarimetric modes are compared to theoretically expected diffraction limited point spread functions. In long slit spectroscopy, wavelength calibration and spectral resolution is demonstrated for the different grism, slit and camera combinations.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Markus Hartung, Peter Bizenberger, Armin Boehm, Werner Laun, Rainer Lenzen, Karl Wagner, "First test results and calibration methods of CONICA as a stand-alone device", Proc. SPIE 4008, Optical and IR Telescope Instrumentation and Detectors, (16 August 2000); doi: 10.1117/12.395543; https://doi.org/10.1117/12.395543