Paper
15 November 2000 Tracking MODIS gain change on-orbit by relating it to SRCA lamp parameters
Harry E. Montgomery, Nianzeng Che, Jeff Bowser
Author Affiliations +
Abstract
This paper addresses the methodology and algorithm for using the Spectro-Radiometric Calibration Assembly (SRCA) lamp electrical parameters to track its output band radiance change. This allows the use of the SRCA to track MODurate resolution Imaging Spectroradiometer (MODIS) detector gain variation at different orbit positions because the MODIS dn change is attributed to its gain change after subtracting the SRCA output radiance change itself. Pre-launch test data show that using lamp current as a parameter is valid. Orbit data prove that the approach is valid for longer periods of time for all Solar Reflective Bands (SRBs). Data indicate that the MODIS has undetectable gain change at different orbit positions.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Harry E. Montgomery, Nianzeng Che, and Jeff Bowser "Tracking MODIS gain change on-orbit by relating it to SRCA lamp parameters", Proc. SPIE 4135, Earth Observing Systems V, (15 November 2000); https://doi.org/10.1117/12.494247
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Lamps

MODIS

Sensors

Temperature metrology

Feedback control

Mirrors

Resistance

Back to Top