Paper
22 September 1983 Low Temperature Infrared Source Calibration And Traceability At Arnold Engineering Development Center (AEDC)
H. R. Little, Jay Hiatt, K A. Lienemann
Author Affiliations +
Abstract
The Aerospace Chamber (7V) at AFDC is a radiometric calibration facility for cold back-ground space-based long-wavelength infrared sensors. A working standard, low-temperature blackbody (BB) has been developed for use in establishing radiometric calibrations that are directly traceable to the National Bureau of Standards (NBS). A description of the BB and the NBS calibration results are presented. This standa.rd source has been utilized to calibrate a phosphorous-doped, silicon bolometer which serves as a transfer device for the calibration of new blackbody sources. The electrically self-calibrating feature of this bolometer has been used to normalize variations in responsivity from one installation to anotter over a period of five years. For infrared (IR) sensor testing, the radiometric quantity of interest is beam irradiance at the sensor aperture. The calibration transfer process which is used to relate the working standard to attenuated sources, is described and the transfer devices are discussed.
© (1983) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
H. R. Little, Jay Hiatt, and K A. Lienemann "Low Temperature Infrared Source Calibration And Traceability At Arnold Engineering Development Center (AEDC)", Proc. SPIE 0416, Applications of Optical Metrology: Techniques and Measurements II, (22 September 1983); https://doi.org/10.1117/12.935929
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Sensors

Bolometers

Black bodies

Infrared radiation

Standards development

Infrared sensors

RELATED CONTENT


Back to Top