9 November 2000 Integrated, multi-angle, low turbidity measurement using fluorescent plastic optical fibre
Author Affiliations +
Proceedings Volume 4185, Fourteenth International Conference on Optical Fiber Sensors; 418556 (2000) https://doi.org/10.1117/12.2302329
Event: Fourteenth International Conference on Optical Fiber Sensors, 2000, Venice, Italy
Abstract
The development of a contact and a non-contact sensor to measure low turbidity of potable water using fluorescent optical fibres over the range 0 to 1 FTU is presented. The contact sensor had a sensitivity of 3.3 nW I unit FTU and an accuracy of ± 0.1 to 0. 15 FTU. The non-contact version of the sensor had a sensitivity of 1.17 nW/ unit FTU.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
R. P. Chandy, "Integrated, multi-angle, low turbidity measurement using fluorescent plastic optical fibre", Proc. SPIE 4185, Fourteenth International Conference on Optical Fiber Sensors, 418556 (9 November 2000); doi: 10.1117/12.2302329; https://doi.org/10.1117/12.2302329