5 October 2000 Calibrated equipment for return loss of fiber optic passive components
Author Affiliations +
Proceedings Volume 4223, Instruments for Optics and Optoelectronic Inspection and Control; (2000) https://doi.org/10.1117/12.401755
Event: Optics and Optoelectronic Inspection and Control: Techniques, Applications, and Instruments, 2000, Beijing, China
Abstract
We report a calibrated equipment for return loss (RL) of fiber optic passive components, the expanded uncertainty for this equipment are 0.80dB(0--60)dB, 1.60dB(-60- -70)dB with 1300 nm and 1550 nm wavelength, RL measurement repeatability to 0.25dB(--60)dB, 0.50dB(-60--70)dB, and stability to 0.20dB(0- -60)dB, 0.40dB(-60--70)dB are obtained by experiments.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Limeng Wang, Limeng Wang, Yongsheng Wu, Yongsheng Wu, Xuhui Luo, Xuhui Luo, } "Calibrated equipment for return loss of fiber optic passive components", Proc. SPIE 4223, Instruments for Optics and Optoelectronic Inspection and Control, (5 October 2000); doi: 10.1117/12.401755; https://doi.org/10.1117/12.401755