2 March 2001 Efficient autoscanned single-frequency cw dye laser
Author Affiliations +
Abstract
We report on a CW single-frequency dye laser with a 1-W threshold power. Output power of the laser working on the Rhodamine 19 dye was 100 mW out of 2.4 W of the argon ion pump. Smooth frequency scanning in a 20-GHz region with up to 20-GHz/s rate was implemented in a linear dye laser cavity. A special resonator auto-adjustment system with internal reference features the short-term radiation linewidth down to 3 MHz/s and long-term stability level of 150 MHz/hour without any external frequency stabilization systems. Explained are arrangements of original execution units included into the laser control system. The control system's features are discussed.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
S. M. Kobtsev, S. M. Kobtsev, A. V. Korablev, A. V. Korablev, S. V. Kukarin, S. V. Kukarin, V. B. Sorokin, V. B. Sorokin, "Efficient autoscanned single-frequency cw dye laser", Proc. SPIE 4353, Laser Optics 2000: Control of Laser Beam Characteristics and Nonlinear Methods for Wavefront Control, (2 March 2001); doi: 10.1117/12.417733; https://doi.org/10.1117/12.417733