31 October 2001 Preliminary results from multicenter clinical trials for detection of cervical squamous intraepithelial lesions using a novel full-field evoked-tissue-flourescence-based imaging instrument
Author Affiliations +
Abstract
We report preliminary results from a multi-center trial of evaluating the performance of a novel, full-field multi-spectral tissue fluorescence imaging system (CerviscanTM) designed to detect cervical squamous intraepithelial lesions. Spectral data from multiple regions of 48 patients enrolled from clinical performance sites in the US and Canada are included in this preliminary analysis. This includes 30 women used for the training set and 18 used for the testing set. In the testing set, 37/43 SIL and 70/80 NonSIL regions were correctly identified for a sensitivity of 86% and specificity of 87.5%. CerviscanTM locates cervical precancerous lesions with high sensitivity and specificity and has the potential to permit 'see and treat' patient management.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Anupam K. Dattamajumdar, D. Wells, Julie R. Parnell, Dipankur Ganguly, Thomas C. Wright, "Preliminary results from multicenter clinical trials for detection of cervical squamous intraepithelial lesions using a novel full-field evoked-tissue-flourescence-based imaging instrument", Proc. SPIE 4432, Diagnostic Optical Spectroscopy in Biomedicine, (31 October 2001); doi: 10.1117/12.447157; https://doi.org/10.1117/12.447157