23 May 2002 Online measurement of urea concentration in spent dialysate during hemodialysis
Jonathon T. Olesberg, Ben Armitage, Mark A. Arnold, Michael Flanigan
Author Affiliations +
Abstract
We describe on-line optical measurements of urea concentration during the regular hemodialysis treatment of several patients. The spectral measurements were performed in the effluent dialysate stream after the dialysis membrane using an FTIR spectrometer equipped with a flow-through cell. Spectra were recorded across the 5000-4000 cm-1 (2.0-2.5 micrometers at 1-minute intervals. Optically determined concentrations matched concentrations obtained from standard chemical assays with a root-mean-square error of 0.29 mM for urea (0.8 mg/dl urea nitrogen), 0.03 mM for creatinine, 0.11 mM for lactate, and 0.22 mM for glucose. The observed concentration ranges were 0-11 mM for urea, 0-0.35 mM for creatinine, 0-0.75 mM for lactate, and 9-12.5 mM for glucose.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Jonathon T. Olesberg, Ben Armitage, Mark A. Arnold, and Michael Flanigan "Online measurement of urea concentration in spent dialysate during hemodialysis", Proc. SPIE 4624, Optical Diagnostics and Sensing of Biological Fluids and Glucose and Cholesterol Monitoring II, (23 May 2002); https://doi.org/10.1117/12.468312
Lens.org Logo
CITATIONS
Cited by 5 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Urea

Calibration

Glucose

Urea concentration

Absorbance

Optical filters

Process modeling

Back to Top