3 May 2002 Development and evaluation of a real-time three-dimensional CT (4D-CT) scanner
Yasuo Saito, Hiroshi Aradate, Hiroaki Miyazaki, Yoji Kudo, Kazuhiko Tsujita, Nobuhisa Shimadu, Yuji Sawanaga
Author Affiliations +
Abstract
4D-CT scanner is a high-speed cone-beam CT with good low- contrast detectability. In a single rotation, a voxel data set with higher spatial resolution over a wide z-axis can be obtained. Using continuous rotation, temporarily continuous voxel data sets and 3D dynamic images can be acquired. We have developed a 4D-CT prototype system using a large area 2D detector with high speed rotating gantry. The key technologies for 4D-CT scanner are a large area 2D detector, high-speed and continuous rotating gantry with high-speed data transfer. The full size 2D detector for prototype system has 912 channels, 256 detector rows for 0.5 mm slice thickness, and the sampling speed is 900 view/s. The rotation speed of the gantry is 1.0 s/rotation, and the speed of the data transfer system must be more than 5Gbps. The data transfer system consists of laser diode and photo diode. We have got real-time volume images of crayfish, moving jaw and breathing lung. We will start clinical research at next stage, and continue to improve the image quality.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yasuo Saito, Hiroshi Aradate, Hiroaki Miyazaki, Yoji Kudo, Kazuhiko Tsujita, Nobuhisa Shimadu, and Yuji Sawanaga "Development and evaluation of a real-time three-dimensional CT (4D-CT) scanner", Proc. SPIE 4682, Medical Imaging 2002: Physics of Medical Imaging, (3 May 2002); https://doi.org/10.1117/12.465631
Lens.org Logo
CITATIONS
Cited by 3 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Computed tomography

Prototyping

3D scanning

3D image processing

Lung

Laser scanners

Back to Top