20 September 2002 Analysis and design of a pulsed-light emitter for time-resolved fluoroimmunoassay
Author Affiliations +
Proceedings Volume 4927, Optical Design and Testing; (2002); doi: 10.1117/12.471680
Event: Photonics Asia, 2002, Shanghai, China
Abstract
This paper deals with a pulsed-light emitter, which is a perfect source for time-resolved fluorescence immunoassay (TRFIA). Using a time-controlled circuit and an energy-controlled circuit, the set can emit high frequency pulsed-light. The frequency is 1000Hz. The wavelength range is from 280nm to 280nm, the pulse width is 54μs, and the output power of ultraviolet radiation is more than 200μw.
© (2002) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zhen Tian, Zhouyi Guo, "Analysis and design of a pulsed-light emitter for time-resolved fluoroimmunoassay", Proc. SPIE 4927, Optical Design and Testing, (20 September 2002); doi: 10.1117/12.471680; https://doi.org/10.1117/12.471680