5 June 2003 Image quality of flat-panel cone beam CT
Georg Rose, Jens Wiegert, Dirk Schaefer, Klaus Fiedler, Norbert Conrads, Jan Timmer, Volker Rasche, Niels Noordhoek, Erhard Klotz, Reiner Koppe
Author Affiliations +
Abstract
We present results on 3D image quality in terms of spatial resolution (MTF) and low contrast detectability, obtained on a flat dynamic X-ray detector (FD) based cone-beam CT (CB-CT) setup. Experiments have been performed on a high precision bench-top system with rotating object table, fixed X-ray tube and 176 x 176 mm2 active detector area (Trixell Pixium 4800). Several objects, including CT performance-, MTF- and pelvis phantoms, have been scanned under various conditions, including a high dose setup in order to explore the 3D performance limits. Under these optimal conditions, the system is capable of resolving less than 1% (~10 HU) contrast in a water background. Within a pelvis phantom, even inserts of muscle and fat equivalent are clearly distinguishable. This also holds for fast acquisitions of up to 40 fps. Focusing on the spatial resolution, we obtain an almost isotropic three-dimensional resolution of up to 30 lp/cm at 10% modulation.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Georg Rose, Jens Wiegert, Dirk Schaefer, Klaus Fiedler, Norbert Conrads, Jan Timmer, Volker Rasche, Niels Noordhoek, Erhard Klotz, and Reiner Koppe "Image quality of flat-panel cone beam CT", Proc. SPIE 5030, Medical Imaging 2003: Physics of Medical Imaging, (5 June 2003); https://doi.org/10.1117/12.479973
Lens.org Logo
CITATIONS
Cited by 14 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Spatial resolution

X-ray computed tomography

Sensors

Image quality

Bone

Modulation transfer functions

X-rays

Back to Top