5 June 2003 Photon-counting energy-sensitive micropattern gas detector for digital radiography
Author Affiliations +
Abstract
A photon counting energy sensitive micro-pattern gaseous detector is being developed for projection radiography. Two cascaded gas electron multipliers (GEMs) are used as gain stages following a drift region that acts as an absorption volume. Active area of the detector is 7cm x 7cm and the signal is read out by 0.2 mm pitch strips. An aluminum pressure vessel was designed as an integral part of the detector, which operates at pressures up to 10 atm. Voltage-gain curves were obtained for Ar and Kr mixtures. Energy resolution for a 5.9 keV Fe-55 was determined to be 20% FWHM. Energy spectra for a conventional x-ray generator operating at various voltages were also obtained. Detector MTF and DQE were calculated using linear systems analysis. The effect of primary electron range, a dominant factor in determining spatial resolution, and the effect of fluorescent photon reabsorption were studied. Operation of the detector for imaging at higher pressures with Kr and Xe mixtures is in progress.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Mustafa C. Altunbas, Eugene Franklin Barasch, Daniel R. Bednarek, and Stephen Rudin "Photon-counting energy-sensitive micropattern gas detector for digital radiography", Proc. SPIE 5030, Medical Imaging 2003: Physics of Medical Imaging, (5 June 2003); https://doi.org/10.1117/12.480001
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

X-rays

Luminescence

Modulation transfer functions

Krypton

Xenon

Absorption

Back to Top