2 June 2003 Portable phase measuring interferometer using Shack-Hartmann method
Author Affiliations +
Abstract
A real-time inspection is useful and effective to optimize lens aberrations of excimer-exposure sytem, which can expose patterns less than 100 nm. We have developed a portable i.e., compact and lightweight phase measuring interferometer (P-PMI), which can be attached to a stage of the exposure system during real-time monitoring the aberration of the projection lens mounted on the exposure system. Measured repeatability of the wavefront measurement is ab out 0.1 mλ and tool-to-tool difference is 0.6mλ. Measured wavefront during adjusting a projection lens agree dwell with a simulated result. LWA was successfully optimized using P-PMI data.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Toru Fujii, Toru Fujii, Jun Kougo, Jun Kougo, Yasushi Mizuno, Yasushi Mizuno, Hiroshi Ooki, Hiroshi Ooki, Masato Hamatani, Masato Hamatani, } "Portable phase measuring interferometer using Shack-Hartmann method", Proc. SPIE 5038, Metrology, Inspection, and Process Control for Microlithography XVII, (2 June 2003); doi: 10.1117/12.482699; https://doi.org/10.1117/12.482699