14 October 2003 Delivery system for laser medical instrument
Author Affiliations +
Abstract
Investigation of the special constructed hollow glass waveguides was realized. Maximum mean power transmitted via this delivery system was 5.8 W (for alexandrite radiation) or 5.1 W (for mid infrared Er.YAG light). Maximum output intensity 173 GW/cm2 was reached for delivery of 55 psec long Nd:YAG pulses.
© (2003) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Helena Jelinkova, Helena Jelinkova, Michal Nemec, Michal Nemec, Jan Sulc, Jan Sulc, Pavel Cerny, Pavel Cerny, Mitsunobu Miyagi, Mitsunobu Miyagi, Yi-Wei Shi, Yi-Wei Shi, Yuji Matsuura, Yuji Matsuura, } "Delivery system for laser medical instrument", Proc. SPIE 5143, Novel Optical Instrumentation for Biomedical Applications, (14 October 2003); doi: 10.1117/12.500450; https://doi.org/10.1117/12.500450