Translator Disclaimer
22 January 2004 Series representation of the instrument function of a two-aperture radiometer
Author Affiliations +
Abstract
Recently it was reported a method to calculate the instrument function of a two-aperture radiometer which describes the energy transport through two apertures using the theory of partial coherence. The result of that work was expressed as a multiplication of a two-fold integral and its complex conjugate. In this work we solve partially this two-fold integral, particularly we introduce a semi-cicle angular integral that reduces the integration. This new representation allows a faster numerical evaluation as well as an easier interpretation of energy transport for radiometric considerations.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Juan Carlos Solorio-Leyva, Jose Guadalupe Suarez-Romero, Juan B. Hurtado-Ramos, Eduardo Tepichin-Rodriguez, and Jose G. Cortes-Reynoso "Series representation of the instrument function of a two-aperture radiometer", Proc. SPIE 5178, Optical Modeling and Performance Predictions, (22 January 2004); https://doi.org/10.1117/12.503965