4 March 2004 On-site comprehensive analysis of explosives using HPLC-UV-PAED
Author Affiliations +
Proceedings Volume 5270, Environmental Monitoring and Remediation III; (2004) https://doi.org/10.1117/12.516186
Event: Optical Technologies for Industrial, Environmental, and Biological Sensing, 2003, Providence, RI, United States
Abstract
High-performance liquid chromatography with ultra violet and photo-assisted electrochemical detection (HPLC-UV-PAED) has been developed for the sensitive and selective detection of explosives in ground water and soil extracts. Fractionation and preconcentration of explosives is accomplished with on-line solid phase extraction (SPE), which minimizes sample pretreatment and enables faster and more accurate on-site assessment of a contaminated site. Detection limits are equivalent or superior (i.e., <1 part-per-trillion for HMX) to those achieved using the Environmental Protection Agency (EPA) Method 8330. This approach is more broadly applicable, as it is capable of determining a wider range of organic nitro compounds. Soil samples are extracted using pressurized fluid extraction (PFE), and this technique is automatable, field-compatible, and environmentally friendly, adding to the overall efficiency of the methodology.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Ronita L. Marple, William R. LaCourse, "On-site comprehensive analysis of explosives using HPLC-UV-PAED", Proc. SPIE 5270, Environmental Monitoring and Remediation III, (4 March 2004); doi: 10.1117/12.516186; https://doi.org/10.1117/12.516186
PROCEEDINGS