Paper
13 July 2004 Ex vivo accommodation simulator II: concept and preliminary results
Author Affiliations +
Abstract
A new instrument was developed to simulate the accommodation of excised human and animal crystalline lenses. The sclera is attached to 8 radially arranged translation stages and dissected into 8 segments. Change in accommodation is simulated by pulling and releasing the scleral segments, varying the load on the zonules and the capsular bag. Each translation stage is equipped with a sensitive force transducer for simultaneous recording of load and displacement. Additionally, changes in optical power, lens diameter, lens thickness and lens shape can be observed and quantified while the lens is stretched and released. The sensitivity and accuracy of the instrument was verified using steel springs and elastic bands. Preliminary results from sensitivity testing and instrument calibration are presented.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Klaus Ehrmann, Arthur Ho, and Jean-Marie Parel "Ex vivo accommodation simulator II: concept and preliminary results", Proc. SPIE 5314, Ophthalmic Technologies XIV, (13 July 2004); https://doi.org/10.1117/12.539111
Lens.org Logo
CITATIONS
Cited by 4 scholarly publications and 19 patents.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Transducers

Eye

Crystals

Motion measurement

Sclera

Calibration

Tissues

Back to Top